Flowmeter
Type
Flow range
Ventilation range
Flow resistance
Accuracy |
Bidirectional digital turbine Ø 18 mm
4 l/sec
0-50 liters/min
<0.7 cm H2O/l/s @ 3 l/s
±2% |
Oxygen Analizer
Type
Accuracy
Response Time
Warm up time
Lifespan |
GVC (Galvanic Fuel cell)
+/- 0.02%
150 msec
Not required
12-18 months |
| HR Monitor |
Polar Receiver & transmitter |
| Pressure sensor |
Range 15-115 Kpa |
| Test |
RMR, Body composition, Daily caloric intake, Weight management programs, Antropometric measurment, CV risk stratification, Par-Q |
| Measured parameters |
RMR (Kcal/day),Daily Caloric Intake (Kcal/day), FeO2, VE, RF, HR, Barometric pressure, FAT%, BMI, Waist circumference, CV risk stratification |
| Predicted values |
American College of Sport Medicine guidelines for testing & exercise prescriptions |
| Power |
Rechargeable Ni-Mh batteries or power supply medical grade approved for main use |
| Inteface |
USB-A (PC), USB-B (printer), RS 232 (treadmill and bike), Polar HR receiver |
| Dimensions |
198x238x76 mm (7.8x9.4x3 in) |
Weight |
1200 g (2.6 lbs) |
| CE Mark |
Fitmate meets the European Directive 93/42 on Medical Device |
| Safety Standards |
Safety Standards: EN 60601-1 (safety)/EN 60601-1-2 (EMC) |
| External printers compatible |
Brother HL-1240, Brother HL-1250, HP laserjet 1200 series, Brother HL-1070, HP laserjet 1010, HP DeskJet 656C |
| Languages Available |
English & Italian |