Specification

Flowmeter RMR
Type
Flow range
Ventilation range
Flow resistance
Accuracy

Bidirectional digital turbine Ø 18 mm
4 l/sec
±2%
<0.7 cm H2O/l/s @ 3 l/s
±2%
Flowmeter VO2 Max
Type
Flow range:
Ventilation range
Flow resistance
Accuracy

Bidirectional digital turbine Ø 28 mm
0.03 - 18 l/sec
10 - 300 liters/min
<0.7 cm H2O/l/s @ 12 l/s
±2%
HR monitor Polar receiver & transmitter

Oxygen Analizer
Type
Accuracy
Response Time
Warm up time
Lifespan


GVC (Galvanic Fuel cell)
+/- 0.02%
150 msec
Not required
12-18 months
Pressure sensor

Range 15-115 Kpa

Test

RMR, Body composition, Daily caloric intake, Weight management programs, Standard measurment, CV risk stratification, Par-Q, Muscolar strenght, Muscular endurance, Flexibility, VO2 max test, VO2 Submax test, Rockpot walking test, 1.5 mile run test, Cooper 12 minute run test, Exercise prescription

Measured parameters RMR (Kcal/day), Daily Caloric Intake (Kcal/day), FeO2, VE, RF, HR, VO2 Max, VO2 max extrapolated, class percentile for muscular strength, muscular endurance , flexibility, barometric pressure, FAT%, BMI, Waist circumference, CV risk stratification
Predicted values

American College of Sport Medicine guidelines for testing & exercise prescriptions

Power Rechargeable Ni-Mh batteries or power supply medical grade approved for main use
Inteface USB-A (PC), USB-B (printer), RS 232 (treadmill and bike), Polar HR receiver
Dimensions 198x238x76 mm (7.8x9.4x3 in)

Weight

1200 g (2.6 lbs)

CE Mark

Fitmate Pro meets the European Directive 93/42 on Medical Device

Safety Standards Safety Standards: EN 60601-1 (safety)/EN 60601-1-2 (EMC)
External printers compatible Brother HL-1240, Brother HL-1250, HP laserjet 1200 series, Brother HL-1070, HP laserjet 1010, HP DeskJet 656C
Languages Available English & Italian
Contact Us | Webmaster | Disclaimer | Home